Flow:600 mL/min;Dewpoint at 7 barg (100 psig):-73ºC (-99.4ºF)
Model: NM PLUS 600 FID STATION
Order NO.:
312
Specific parameters will be adjusted according to product upgrades, without notice.
DESCRIPTION
The Chemtron Gas NM Plus FID Station is a unique instrument that combines the reliability of the hydrogen generator with a zero air generator into one compact package.
The FID Station can be installed under the GC taking up no additional bench space. This simple but effective instrument can supply all your carrier gas and FID gas requirements. Designed as a hazard free alternative to high pressure cylinders, all that is required is deionized water, compressed air and a standard electrical supply for weeks of continuous operation.
Innovative software control allows unrivaled operational performance and safety as well as the additional options of auto water feed, remote networking and cascading for built in redundancy. With a maximum output capacity of 1350 mL/min, one generator can supply up to 32 GCs. The compact design allows the generator to be installed directly in the laboratory eliminating the requirement for long gas lines and guaranteeing the delivery of high purity gas to your GC.
A sophisticated control system connected to an easy to use touch screen continuously monitors vital operating parameters to ensure a safe and consistent performance.
Built in sensors will shut the generator down if internal/external leaks are present, contaminated water, low water or over pressure. This is why Chemtron Gas generator meet the strict safety guidelines to be certified for CE, FCC, and MET (UL and CSA compliant).
INCREASE EFFICIENCY
A constant gas supply with a guaranteed purity, eliminates interruptions of analyses to change cylinders and reduces the amount of instrument re-calibration required.
ENHANCE RESULTS
Hydrogen as a carrier gas is faster and more sensitive than expensive helium, with run time savings of 25% to 35% without a decline in
resolution. The use of hydrogen as a carrier gas allows lower temperature elution, thus extending the life of the chromatograph column.
IMPROVE SAFETY
Gas is produced on demand, which allows for the safe use of the hydrogen generator when cylinders are prohibited or regarded as potentially dangerous. Sophisticated software control and full alarm capability, including for hydrogen leaks, gives the user full control of the gas supply.
ENHANCE PERFORMANCE
Gas generator can be installed in the laboratory close to the instrument, eliminating the need for long gas lines from external cylinder
supplies. A constant guaranteed high purity gas supply improves stability and ensures greater reproducibility of results.
FEATURES
Produces a continuous supply of hydrogen | Compact and robust design to accommodate up to 80 kg (176 lbs.) top loading - suits all GC
manufactures | Space saving design, the GC sits directly on top of the generator | On-demand supply 24/7 | H2 Flow rate: 100 to 1350
mL/min & zero air up to 5 L/min | H2 Purity: >99.99996% & zero air <0.1 ppm THC | Pressure: 11 barg (160 psig) | Proprietary 100%
titanium cell technology | Unique permeation membrane drying system | Cold dual dynamic regeneration dryer | USB connectivity | 2-year complete cell and product warranty | Easy to install, operate and maintain
BENEFITS
Eliminates dangerous high pressure cylinders helping to keep your employees safer | Frees up bench space | Removes the logistics,
inconvenience, downtime and costs of a cylinder system | Flow capacity and purity to match your specific instrument demands | Ideal
for all GC and GC/MS applications | Meets and exceeds the requirements for the most demanding GC applications | Superior hydrogen
production with reliable long life cell | Minimal maintenance - low cost of ownership | PC Monitoring for maintenance, diagnostics and
remote connection | Peace of mind | Improve your laboratory work flow and productivity
OPERATING DIAGRAM
Hydrogen is produced from the hydrolysis of deionized water across a PEM (proton exchange membrane),housed in a 100% titanium cell. The ouput hydrogen is dried via a dual stage process, a gas liquid separator and a unique cold static automatic dryer.
In addition to water all that the generator requires is a standard connection and supply of electricity for a continuous 24/7 supply of high purity hydrogen. Consumable items are limited to the replacement of a deionizer bag every six months.
Specifications
Model |
NM PLUS 600 FID STATION |
Flow mL/min |
600 |
Purity |
>99.99996% |
Dewpoint at 7 barg (100 psig) |
-73ºC (-99.4ºF) |
Outlet pressure barg (psig) |
1.4 to 11 (20 to 160) |
Technology |
PEM (proton exchange membrane) - 100% titanium cell |
Drying system |
No maintenance cold dual dynamic regeneration system |
Deionized water quality |
Minimum <1 micro S/cm @ 25ºC - 1 Mohm-cm @ 25ºC - ASTM II Recommended <0.2 microS/cm @ 25ºC - 5 Mohm-cm @ 25ºC - ASTM II |
Internal water tank (liters) |
7 |
Safety |
Automatic shut down - internal/external hydrogen leak, overpressure and low water |
Display |
Touch screen with operating parameters, system status and safety alarms |
LED Indicators |
Power on/off, system ready, errors |
Interface |
USB mod A |
Electrical supply |
110-120V 60Hz / 220-240V 50 Hz |
Power consumption (watts) |
300 |
Dimensions mm (in) |
690W x 150H x 680D (27W x 6H x 26.7D) |
Weight kg (lbs) |
43 (95) |
Shipping dimensions mm (in) |
890W x 385H x 800D (35W x 15.1H x 31.4D) |
Shipping weight kg (lbs) |
49 (108) |
Operating temp ℃ ( ℉ ) |
15 to 35 (59 to 95) |
Outlet connection |
1/8" Compression |
Certification |
CE, FCC, MET (UL and CSA compliant) |